ECG Setup Menu
Click on the HR parameter box to access the ECG setup menu. Items and settings are
described in the following table.
Quick Reference Table
–
ECG Setup
Menu
Selection
Description
Available Settings
Show All
Leads
Displays all active ECG leads
NOTES
:
• While the
Show All Leads
screen is displayed, other
parameter boxes remain
visible, and alarms and
recordings continue to operate
normally. However, you cannot
use the rotary knob to access
other menus.
• You can also access the
Show
All Leads
screen using the
fixed keys
Fast Access
on the
monitor or the
All ECG
button
on the remote keypad.
• Click on
Show All Leads
to display all
connected available ECG leads.
• Click on
Report
at the bottom of the screen to
print an ECG report on a network laser printer.
• Click on
Notes
at the bottom of the screen to
display remarks about the patient’s
physiological condition. Scroll to the
appropriate note and click again. Notes are
displayed on the screen and can printed on
reports.
• Press the
Main Screen
fixed key to return to
the main screen.
The
Size ECG
Submenu
This submenu allows you to configure the following functions.
Size ALL
ECG
Changes display amplitude of
ECG waveforms
NOTES:
• The monitor's normal QRS
detection threshold is
approximately 0.35 mV
depending on the QRS width.
This threshold is used for
channel size selections of 1, 2,
4, or 8 V/cm (see page 8-16
for more details).
• A channel size selection of
0.25 or 0.5 mV/cm, decreases
the QRS detection threshold
to approximately 0.2 mV. In
this case, these low amplitude
QRS complexes may be
included in the heart rate
calculation if the QRS width is
between 70 and 120 ms.
Click on
Size ALL ECG
to change the amplitude
of all waveforms on the
Show All Leads
and
the main screen.
Channel
<#>
size
Click on
Channel 1 size, Channel 2 size,
or
Channel 3 size
to change the size of the
individual ECG channels.
NOTE
: If you connect a 3-lead cable to the
monitor, the selections
Channel 2 size
and
Channel 3 size
are ghosted.
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...