SvO2/CCO Monitors
An SvO2/CCO monitor measures cardiac output and related hemodynamic
parameters. This Dräger monitor accepts data via an MIB interface from an Abbott Q2
(Q2+) SvO2/CCO monitor or from the Edwards/Baxter Vigilance and Vigilance II
SvO2/CCO monitors.
The following table lists the parameters available for display.
NOTE:
The bedside monitor calculates the Body Surface Area (BSA) using the
patient height and weight that was entered in the “Admit Patient” menu (see page 4-2). It
uses this BSA value to calculate the Index parameters (for example, ICI, CCI, SVRI etc)
for the Baxter MIB device. Although the Baxter device can calculate BSA internally and
send ICI and CCI values to the monitor, the monitor will still calculate the index values
based on its BSA.
Label
Parameters
Units
SvO2/CCO Monitor from
BT*
Blood Temperature
°C
Q2, Vigilance and
Vigilance II
CCO
Continuous Cardiac
Output
L/min.
Q2, Vigilance and
Vigilance II
CCI
CCO Index
L/min./m
2
Vigilance and Vigilance II
ICO
ntermittent Cardiac
Output
L/min.
Q2, Vigilance and
Vigilance II
ICI
ICO Index
L/min./m
2
Vigilance and Vigilance II
SaO2*
Arterial Oxygen
Saturation
%
Vigilance and Vigilance II
SvO2
Venous Oxygen
Saturation
%
Q2, Vigilance and
Vigilance II
SVR
Systemic Vascular
Resistance
dynes x sec x cm
-5
Vigilance and Vigilance II
SVRI
SVR Index
dynes x sec x cm
-5
x m
2
Vigilance and Vigilance II
VO2
O
2
Consumption
mL/min
Vigilance and Vigilance II
DO2
O
2
Delivery
mL/min
Vigilance and Vigilance II
SV
Stroke Volume
mL
Vigilance II
SVI
SV
Index
mL/m
2
Vigilance II
EDV
End Diastolic Volume
mL
Vigilance II
EDVI
EDV Index
mL/m
2
Vigilance II
ESV
End Systolic Volume
mL
Vigilance II
ESVI
ESV Index
mL/m
2
Vigilance II
EF
Ejection Fraction
%
Vigilance II
The Abbott Oxymetrix 3 SvO2/CCO monitor also provides CCO and SvO2. See the documentation
accompanying that device for information about these parameters and their calculation.
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...