![Dräger Infinity Delta Скачать руководство пользователя страница 450](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779450.webp)
26-12
D
ELTA
/D
ELTA
XL/K
APPA
VF8
26 P
ULSE
C
ONTOUR
C
ARDIAC
O
UTPUT
(P
I
CCO)
MONITORING
16. Inject the bolus solution into the patient’s bloodstream only after you see the
READY
message. A thermodilution curve appears, displaying the change in
blood temperature and a p-CO value is computed.
17. Repeat step 15 to take an additional measurement, making sure you wait for
the
READY
message. If a temperature drop is not detected within four
minutes, the Averaging screen closes, and the current p-CO values will be
averaged and saved. You must repeat steps 15 and 16 if additional p-CO
measurements are necessary.
Averaging p-CO Measurements
Differences in injection technique can cause variations in measurements performed on
the same patient. To compensate for such discrepancies, you can review results of up to
five measurements and use them to compute a p-CO average. The p-CO Averaging
screen is displayed whenever you begin a p-CO measurement. The Review Curves
screen duplicates the five values displayed on the p-CO Averaging screen with their
corresponding thermodilution curves.
NOTE:
If the
READY
message fails to appear or appears only intermittently, the blood
temperature may be unstable. Check that all connections are correct and that the blood
temperature is stable and valid and repeat step 15 (see page 26-23 for suggestions to
optimize PiCCO measurements).
NOTE:
During measurements, the patient's condition needs to be stable and
administration of infusions and injections should be discontinued. The blood
temperature (p-BT) should not be lower than 30 °C.
Iced (< 8°C) or room-temperature injectate < 24 °C can be used. The maximum
temperature decrease (
Δ
T) at the measurement point should be > 0.15 °C.
To achieve a usable thermodilution signal, Pulsion recommends that the volume
of injectate to be used should be based on the following table:
EVLWI < 10
EVLWI > 10
EVLWI < 10
kg body weight iced
iced
room temperature
< 3
2 ml
2 ml
3 ml
< 10
2 ml
3 ml
3 ml
< 25
3 ml
5 ml
5 ml
< 50
5 ml
10 ml
10 ml
< 100
10 ml
15 ml
15 ml
> or = 100
15 ml
20 ml
20 ml
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...