CAUTION:
If the PiCCO pod appears to be damaged, do not use the pod; contact Dräger
instead.
If the pulse contour measurement values seem unreasonable, perform a
thermodilution measurement. This will automatically recalibrate the PiCCO
pod and provide more accurate continuous pulse contour cardiac output
measurements. Recalibration is recommended if the hemodynamic conditions
change significantly, due to volume shifts or medication changes.
Inaccurate measurements can be caused by incorrectly placed catheters,
defective connections or sensors, and by electromagnetic interference (for
example, electric blankets, electric coagulation).
The displayed GEDV value may be erroneously high in patients with an aortic
aneurysm if thermodilution measurements are made via the femoral artery.
Aortic aneurysms may cause the displayed blood volume (GEDV/ITBV) derived
by thermodilution measurements to be erroneously high.
As the pulse contour cardiac output of children has not been sufficiently
validated thus far, the CO should be checked by thermodilution before
therapeutic interventions. Recalibration is recommended with significant
changes in hemodynamic conditions, such as volume shifts or changes to
medication.
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...