Kappa CPU Base Unit
Physical Attributes
Size (H x W x D):
102 x 368 x 368 mm (4 x 14.5 x 14.5 in)
Weight:
8.4 kg (19 lb.)
Cooling:
Fan
Enclosure:
Plastics: ABS/PC, FR 110
Painted, cold-rolled steel
Printed circuit boards:
Board: Glass/epoxy Fr4
Solder: Lead/tin
Copper etch
Lithium battery
Heatsink assemblies:
Cast aluminum
NBP assembly:
Silicon tubing, steel, copper wire
Electrical Specifications
Power requirements:
100-240 VAC, 2.5 A to1.3 A 50/60 Hz
Protection class:
Internally powered (per IEC 60601-1) and for use with specified
Class 1 power supplies.
Chassis leakage:
300
μ
A @ 120 VAC
500
μ
A @ 240 VAC
Battery type:
Lithium Ion
Battery charge duration:
20 minutes
Battery charging time:
3.5 hours at 25 ºC
Patient leakage current:
≤
10
μ
A
Mode of operation:
Continuous with external power supply; for a limited time with
battery backup
NOTES:
• Battery life varies with parameter configuration. The battery life specified above is under the
following load conditions: MultiMed
with SpO
2
sensor, 2 temperature probes, HemoMed pod with
4 IBP transducers and a catheter, NBP taking measurements every 15 minutes, LCD Transport
Brightness at 50%, and no continuous tone being generated.
• Advanced Communication Options are not supported during battery operation.
Environmental Requirements
Temperature range:
Operating: 10 to 45 °C (50 to 113 °F)
Storage:
–15 to +50 °C (–5 to +122 °F)
Relative humidity:
Operating: 20 to 90 %, non-condensing
Storage:
10 to 95 % (with packaging)
Atmospheric pressure:
Operating: 525 to 795 mmHg (70 to 106 kPa)
Storage:
375 to 795 mmHg (50 to 106 kPa)
Protection against ingress of
water:
IPX1
Analog Output
Signals:
ECG, arterial blood pressure (ART)
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...