![Dräger Infinity Delta Скачать руководство пользователя страница 460](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779460.webp)
26-22
D
ELTA
/D
ELTA
XL/K
APPA
VF8
26 P
ULSE
C
ONTOUR
C
ARDIAC
O
UTPUT
(P
I
CCO)
MONITORING
CVP Value
Allows user to select a CVP value.
NOTE
:
Auto
CVP Value
, if available,
is used until the manual
CVP Value
menu
is enabled. If auto
CVP Value
is unavailable, a default of 5mmHg is
used until manual
CVP Value
is
menu
is enabled.
5 mmHg
(default)
0 mmHg
to
300 mmHg
(in increments of 1 mmHg)
Injectate Volume
Selects the injectate volume.
15.0 cc
(default),
2.0 cc
to
20 cc
, in increments of
1.0 cc
p-CO Start
Brings up a p-CO Averaging screen
and starts p-CO measurement.
NOTES
:
• After each thermodilution
measurement completes, select
p-CO Start
in the Averaging screen
to perform another measurement
(the monitor does not automatically
start a new measurement).
• A p-CO measurement can be
started using the
CO Start
key on
the PiCCO pod (the keypad's C.O.
key does not work).
• If p-BT or p-IT values are invalid or
out of range, the
p-CO Start
button
is ghosted and the Averaging
screen cannot be accessed. During
this time the
C.O. Start
fixed key on
the PiCCO pod is also inactive.
Not applicable
p-BT
<value>
p-IT
<value>
Read only display of p-BT and p-IT
values.
NOTE
:
If p-BT or p-IT values are out
of range, the values will display “***”.
p-BT range: 25 to 43°C (77 to 109°F)
p-IT range: -3 to 31°C (27 to 88°F)
Not applicable
PBW
<value>
PBSA
<value>
Read only display of PBW and PBSA
values.
Not applicable
Manual CVP Log
CVP value is logged in with a time
stamp.
Not applicable
NOTE
:
A total of 20 entries per
patient can be stored in the log.
Thermodilution Setup Menu
Menu Item
Description
Available Settings
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...