14-10
D
ELTA
/D
ELTA
XL/K
APPA
VF8
14 I
NVASIVE
B
LOOD
P
RESSURE
IBP Y-Cables
The IBP Y-cable increases the monitoring capabilities of the Hemo2 or Hemo4 pod or
allows two IBP parameters to be measured without a hemodynamic pod. When
plugged into the monitor, the Y-cable can accommodate up to two transducers, letting
you take two IBP measurements simultaneously. The monitor automatically assigns
P1a-d, P2a-d and P3a-d as temporary pressure labels when pressure transducers are
initially connected. These temporary pressure labels must be relabeled with an
appropriate permanent label corresponding to the pressure site being monitored.
To connect the IBP Y-cable to the monitor
1. Plug a transducer into the intermediate cable as shown.
2. Plug the other end of the intermediate cable into the Y-cable (7 or 10 pin).
3. Repeat steps 1 and 2 for a second transducer.
4. Plug Y-cable into a hemodynamic pod or into the monitor.
NOTE:
For a complete list of Dräger IBP accessories available with this product, see
page C-19.
1
IBP Y-cable
3
IBP transducers
2
Plug into hemomed connector on
monitor
4
Intermediate cables
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...