Status Messages
Archive
Determines whether the selected event is
stored, recorded, or both. You can view
stored events on the Event Recall screen
(see page 1-27).
NOTE
: The archive function for ASY
(Asystole) and VF (Ventricular Fibrillation)
cannot be disabled.
•
Store
; stores the selected event
•
Record
; generates an alarm recording
of selected event
•
Str./Rec.;
stores an event and
generates an alarm recording (even
when event alarm is turned off)
• OFF
Message
Event Definition
Suggested Action
PVC/min > UL
The PVC/min value is above the upper alarm
limit.
Check the patient.
Cannot Learn
Lead <X>
At the end of the Learning phase, the
dominant normal complex could not be
determined for Lead <X> (one of the two
leads selected for QRS processing).
Arrhythmia analysis proceeds using the other
lead as a source. Lead <X> is ignored until a
Relearn
is initiated.
• Check the leads.
• Choose another lead for QRS
processing.
Relearning
Monitor is learning a normal QRS complex as
a reference template.
Not applicable
Unable to
Learn
After 100 beats, monitor cannot determine
the dominant normal complex on any lead
selected for QRS processing. Learning
continues.
Check the electrode preparation.
Baseline
Artifact
Artifact is blocking arrhythmia classification.
Check the electrode preparation.
Function
Description
Available Settings
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...