You should zero a transducer under the following conditions:
Immediately after you introduce the catheter into the patient’s vascular
system
After you initially connect the transducer to a pressure pod
Before each monitoring session
Before you enter a calibration factor
Whenever you change the tubing or transducer dome
When the message
<IBP> Zero Required
is displayed
The following table outlines zeroing procedures:
Calibration Procedures
Calibration procedures differ depending on whether you are using a disposable
(single-use) or reusable transducer. You do not need to calibrate disposable
transducers, which are already calibrated at the factory to the monitor’s default value
of 100. Prolonged use of reusable transducers, however, may adversely affect
accuracy.
Single Transducer Zero
Simultaneous “Smart Zero”
1. Make sure the transducer is at heart level. Dräger recommends securing the transducer holders
on the front of the hemodynamic pod for proper height.
2. Close the transducer stopcock to the patient and open it to air.
3(a). Click on the parameter box associated with
the transducer you want to zero. The
parameter setup menu appears.
NOTE
: You can also access the parameter
menu as follows:
1) Press the
Menu
fixed key to display the main
menu.
2) Click on
Patient Setup.
3) Click on
Parameters.
4) Scroll to the desired parameter and click.
WARNING: Do not use the ‘Smart Zero’
function if any pressure waveform is flat
(nearly static). Only use the ‘Smart Zero’
function when all the stopcocks are opened
to air.
3(b). Press the
key on the hemodynamic
pod/MPod you wish to zero. The monitor
determines which of the pod’s transducers are
open to air and zeroes them.
NOTE
: If you use step 3(b) and are unable to
zero a particular IBP with the
key, use the
associated parameter box as described in 3(a).
This method can be more effective.
4. Click on
Zero
.
NOTE
: If the procedure is successful, the monitor displays the message:
<IBP> zero accepted
. If
the procedure fails, the monitor displays the message:
<IBP> did not zero.
Check the waveform. If
spikes exceed three millimeters, repeat the procedure. If procedure fails after two attempts,
replace the transducer or consult your hospital’s technical personnel.
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...