REV 1.0
WRM-10P S2 USER’S MANUAL
71
4.2
Converting Resistance Measurements
The WRM can convert the resistance reading of the device under test at its present
temperature to the equivalent resistance value at a different temperature. The conversion is
accomplished by the following formula:
𝑹𝒔 =
𝑹𝒎 (𝑻𝒔 + 𝑻𝒌)
(𝑻𝒎 + 𝑻𝒌)
Where:
Rs
is the resistance at desired temperature
Rm
is the measured resistance
Ts
is the desired reference temperature
Tm
is the temperature at which the resistance was measured
Tk
is the winding material constant
Tk
= 234.5 for copper
Tk
= 240.0 for aluminum
Содержание Vanguard Instruments WRM-10P S2
Страница 28: ...WRM 10P S2 USER S MANUAL REV 1 0 24 Figure 8 Typical V1 and V2 Test Report Printout with Multiple Readings ...
Страница 35: ...REV 1 0 WRM 10P S2 USER S MANUAL 31 Figure 9 Typical LTC Voltage Regulator Test Report Printout ...
Страница 47: ...REV 1 0 WRM 10P S2 USER S MANUAL 43 Figure 11 Typical Transient Test Report ...
Страница 65: ...REV 1 0 WRM 10P S2 USER S MANUAL 61 Figure 12 Typical Internal Test Record Directory Printout ...
Страница 67: ...REV 1 0 WRM 10P S2 USER S MANUAL 63 Figure 13 Typical USB Flash Drive Test Record Directory Printout ...