WRM-10P S2 USER’S MANUAL REV 1.0
32
3.8
Performing a Special Resistance Test
The Special Resistance Test is used to conduct a resistance test for a pre-defined period ranging
from 1 to 45 minutes. The resistance data is recorded at pre-defined intervals (from 10 seconds
to 1 minute). Follow the steps below to perform a Special Resistance Test:
a.
Start from the "START-UP" menu:
Press the
[1]
key (
RUN TEST
).
b.
The following screen will be displayed:
Press the
[3]
key (
SPECIAL TEST V1, V2
) if you would like to perform a V1 and V2 Special
Test.
Press the
[4]
key (
SPECIAL TEST V1
) if you would like to perform a V1 only Special Test.
For this example, we will press the
[3]
key to perform a Special V1, V2 Test.
c.
The following screen will be displayed:
Select the special test interval by pressing the corresponding key on the keypad.
For this example, press the
[3]
key to select 20 seconds. Resistance data will be
recorded at 20 second intervals.
SPECIAL TEST INTERVAL
1. 10S (up to 10 Min)
2. 15S (up to 15 Min)
3. 20S (up to 20 Min)
4. 30S (up to 30 Min)
5. 60S (up to 60 Min)
1.V1 & V2 TEST
2.V1 ONLY TEST
3.SPECIAL TEST V1, V2
4.SPECIAL TEST V1
5.TRANSIENT TEST
1.RUN TEST
2.TEST RECORDS
3.DIAGNOSTICS
4.SETUP
Содержание Vanguard Instruments WRM-10P S2
Страница 28: ...WRM 10P S2 USER S MANUAL REV 1 0 24 Figure 8 Typical V1 and V2 Test Report Printout with Multiple Readings ...
Страница 35: ...REV 1 0 WRM 10P S2 USER S MANUAL 31 Figure 9 Typical LTC Voltage Regulator Test Report Printout ...
Страница 47: ...REV 1 0 WRM 10P S2 USER S MANUAL 43 Figure 11 Typical Transient Test Report ...
Страница 65: ...REV 1 0 WRM 10P S2 USER S MANUAL 61 Figure 12 Typical Internal Test Record Directory Printout ...
Страница 67: ...REV 1 0 WRM 10P S2 USER S MANUAL 63 Figure 13 Typical USB Flash Drive Test Record Directory Printout ...