
REV 1.0
WRM-10P S2 USER’S MANUAL
33
d.
The following screen will be displayed:
Enter the total special test time in minutes (from 1 to 20 minutes) using the keypad and
press the
[ENTER]
key. For this example, we will press the
[1]
key (for 1 minute) and
then press the
[ENTER]
key.
e.
The following screen will be displayed:
1.
YES
Press the
[1]
key (
YES
) if you wish to convert readings to standard temperature.
The following screen will be displayed:
Press the
[1]
key (
COPPER, Tk=234.5
) to select copper as the winding material.
Continue to step f
.
Press the
[2]
key (
ALUMINUM, Tk=225
) to select aluminum as the winding
material.
Continue to step f.
Press the
[3]
key (
MANUALLY ENTER Tk
) to manually enter the Tk value. The
following screen will be displayed:
WINDING MATERIAL
1.COPPER, T
K
=234.5
2.ALUMINUM, T
K
=225
3.MANUALLY ENTER T
K
CONVERT READINGS TO
STANDARD TEMP?
1. YES
2. NO
ENTER SPECIAL TEST
TIME IN MINUTES
(1-20): 0
Содержание Vanguard Instruments WRM-10P S2
Страница 28: ...WRM 10P S2 USER S MANUAL REV 1 0 24 Figure 8 Typical V1 and V2 Test Report Printout with Multiple Readings ...
Страница 35: ...REV 1 0 WRM 10P S2 USER S MANUAL 31 Figure 9 Typical LTC Voltage Regulator Test Report Printout ...
Страница 47: ...REV 1 0 WRM 10P S2 USER S MANUAL 43 Figure 11 Typical Transient Test Report ...
Страница 65: ...REV 1 0 WRM 10P S2 USER S MANUAL 61 Figure 12 Typical Internal Test Record Directory Printout ...
Страница 67: ...REV 1 0 WRM 10P S2 USER S MANUAL 63 Figure 13 Typical USB Flash Drive Test Record Directory Printout ...