WRM-10P S2 USER’S MANUAL REV 1.0
i
SAFETY SUMMARY
FOLLOW EXACT OPERATING PROCEDURES
Any deviation from the procedures described in this operator’s manual may create one or more safety hazards,
damage the WRM-10P S2, or cause errors in the test results. Vanguard Instruments assumes no liability for unsafe
or improper use of the WRM-10P S2. The following safety precautions must be observed during all phases of test
set up, test hookups, testing, and test-lead disconnects.
SAFETY WARNINGS AND CAUTIONS
This device must be used only by
trained operators
. All transformers under test must be
off line and fully isolated
.
SERVICE AND REPAIR
•
Do not install substitute parts or perform any unauthorized modification to any WRM-10P S2 test unit.
•
Repairs must be performed only by Vanguard Instruments factory personnel or by an authorized repair service
provider. Unauthorized modifications can cause safety hazards and will void the manufacturer’s warranty.
EQUIPMENT RATINGS
IP Rating:
The enclosure for the WRM-10P S2 has an IP rating of 32.
Pollution Rating:
The WRM-10P S2 has a pollution rating of 2.
Operating Voltage:
The WRM-10P S2 is rated for use with an operating voltage of 120V or 240V, auto-ranging
±
10% of selected voltage.
Power Cord
: The WRM-10P S2 is supplied with a 16 AWG, 16A power cord with a NEMA 5-15P plug. Replacement
cable must have the same or better rating and is available through the manufacturer.
VENTILATION REQUIREMENTS
The WRM-10P S2 must be operated with the enclosure lid open.
SAFETY SYMBOLS
Indicates that caution should be exercised
OR
Indicates location of chassis ground terminal
CLEANING
To clean the WRM-10P S2:
•
Disconnect all cables and turn the unit off.
•
Use a soft, lint-free cloth to wipe all surfaces clean.
•
Avoid getting moisture in openings and connectors.
•
DO NOT use any cleaning products or compressed air
Содержание Vanguard Instruments WRM-10P S2
Страница 28: ...WRM 10P S2 USER S MANUAL REV 1 0 24 Figure 8 Typical V1 and V2 Test Report Printout with Multiple Readings ...
Страница 35: ...REV 1 0 WRM 10P S2 USER S MANUAL 31 Figure 9 Typical LTC Voltage Regulator Test Report Printout ...
Страница 47: ...REV 1 0 WRM 10P S2 USER S MANUAL 43 Figure 11 Typical Transient Test Report ...
Страница 65: ...REV 1 0 WRM 10P S2 USER S MANUAL 61 Figure 12 Typical Internal Test Record Directory Printout ...
Страница 67: ...REV 1 0 WRM 10P S2 USER S MANUAL 63 Figure 13 Typical USB Flash Drive Test Record Directory Printout ...