vPad-A1 OPERATORS MANUAL
Also on the
CO Adapter
box is a larger, four-pin connector for the typical
Injectate
temperature
probe. To determine if your device connection requires an additional adapter, contact your Datrend
dealer for assistance.
Note
: Since it is not possible to display a selector or button for every
simulation parameter at the same time, the
Cardiac Output
menu
displays, in yellow, a summary list of all the parameter settings which
are currently in use. See
Figure 51
for an example of a parameters
summary.
4.4.2.3 Cardiac Output – Baseline Temperature
Baseline Temperature
sets the temperature of the patient's blood prior to injection of the saline. It
can be set to 36, 37, or 38
E
C.
4.4.2.4 Cardiac Output – Injectate Temperature
Injectate Temperature
sets the temperature of the saline to be injected. The
third dropdown menu presents the list of options as shown in
Figure 52
.
The second-to-last option (‘Spacelabs 20
E
C’) is adjustable by means of user-
accessible potentiometers inside the vPad-PS module. Default resistance for
the Spacelabs
Injectate Temperature
is set at the factory to 68.0K ohm, but
users can change this via the 100K ohm and 5.0K ohm potentiometers as
indicated in
Figure 53
.
Note
: The
Cardiac Output Coefficient
is 0.542 for
0 and 2
E
C, and 0.595 for 20 and 24
E
C.
Figure 51
Figure 52
Figure 53
Manual Operation/Chapter 4
#
Page 44
Содержание vPad-A1
Страница 1: ...Operating Manual MN 103c vPad A1 All in One Patient Simulator and Performance Tester ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 14: ...vPad A1 OPERATORS MANUAL Definitions Page x ...
Страница 28: ...vPad A1 OPERATORS MANUAL Specifications Chapter 1 Page 14 ...
Страница 32: ...vPad A1 OPERATORS MANUAL Overview Chapter 2 Page 18 ...
Страница 102: ...vPad A1 OPERATORS MANUAL Automation and Test Records Chapter 5 Page 88 ...
Страница 104: ...vPad A1 OPERATORS MANUAL Maintenance Chapter 6 Page 90 ...
Страница 107: ...vPad A1 OPERATORS MANUAL Appendix A Page 93 ...
Страница 108: ......