vPad-A1 OPERATORS MANUAL
4.5 Invasive Blood Pressure
4.5.1 General
The
Invasive Blood Pressure
(
IBP
) simulation provides static pressure settings or dynamic output
waveforms suitable for testing up to two invasive blood pressure channels of a multi-parameter monitor.
To access the IBP simulation, touch the
IBP
tab on the left side of the vPad-A1
Main
screen (
Figure 17
).
On selection, the simulation setup menu will be displayed, activating the individual parameter values shown
on the screen.
4.5.2 Parameters
Parameters related to
IBP
can be selected from the dropdown lists at the top of the
screen. There are three modes of operation:
Static Blood Pressure
,
Dynamic Blood
Pressure
, and
Intra-Uterine Pressure
. Touching the top-most menu box will
present a list of the options available, as shown in
Figure 55
.
Selecting any of these options will then display a list of relevant parameters for the
second list. Users may set parameters as described in section
4.1.3 -
Parameter
Values and Options
.
Note
: Since it is not possible to display a selector or button for every simulation
parameter at the same time, the
IBP
menu displays, in yellow, a summary list
of all the parameter settings which are currently in use. See
Figure 56
for an
example of a parameters summary.
Figure 55
Figure 56
Manual Operation/Chapter 4
#
Page 46
Содержание vPad-A1
Страница 1: ...Operating Manual MN 103c vPad A1 All in One Patient Simulator and Performance Tester ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 14: ...vPad A1 OPERATORS MANUAL Definitions Page x ...
Страница 28: ...vPad A1 OPERATORS MANUAL Specifications Chapter 1 Page 14 ...
Страница 32: ...vPad A1 OPERATORS MANUAL Overview Chapter 2 Page 18 ...
Страница 102: ...vPad A1 OPERATORS MANUAL Automation and Test Records Chapter 5 Page 88 ...
Страница 104: ...vPad A1 OPERATORS MANUAL Maintenance Chapter 6 Page 90 ...
Страница 107: ...vPad A1 OPERATORS MANUAL Appendix A Page 93 ...
Страница 108: ......