vPad-A1 OPERATORS MANUAL
4.2.5 Pacemaker Pulses
The
Pacemaker Pulse
simulation tests a patient monitor's ability to correctly display
and/or classify rhythms which include cardiac beats accompanied by artificial
pacing pulses having significantly greater amplitude than a normal QRS complex.
The
Rate
of
Pacemaker Rhythm
simulations is fixed, as the frequency is an
important characteristic of the chosen waveform. The simulations permit
adjustment of both the QRS
Amplitude
and baseline
Artifact
.
Pacer simulation
Mode
is adjustable between Atrial and Ventricular, which affects
the timing of the pacing pulse that is inserted into the paced waveform.
Besides the common ECG parameters,
Pacemaker Pulse
has several additional
settings that are accessible through the second dropdown list (
Figure 36
).
Figure 36
Pacer Rhythm controls the timing of the paced waveform. See the tables which follow for waveform
descriptions. Additional options which are available in
Atrial Mode
are shown in
Figure 37
.
Ventricular
Mode
options are shown in
Figure 38
.
Figure 37
Figure 38
Pacemaker Waveform / Rhythm
Description
Atrial 80 BPM
Atrial pacemaker wave at 80 BPM, with a pacer pulse at the start of each P wave.
Figure 35
Manual Operation/Chapter 4
#
Page 36
Содержание vPad-A1
Страница 1: ...Operating Manual MN 103c vPad A1 All in One Patient Simulator and Performance Tester ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 14: ...vPad A1 OPERATORS MANUAL Definitions Page x ...
Страница 28: ...vPad A1 OPERATORS MANUAL Specifications Chapter 1 Page 14 ...
Страница 32: ...vPad A1 OPERATORS MANUAL Overview Chapter 2 Page 18 ...
Страница 102: ...vPad A1 OPERATORS MANUAL Automation and Test Records Chapter 5 Page 88 ...
Страница 104: ...vPad A1 OPERATORS MANUAL Maintenance Chapter 6 Page 90 ...
Страница 107: ...vPad A1 OPERATORS MANUAL Appendix A Page 93 ...
Страница 108: ......