vPad-A1 OPERATORS MANUAL
• Simulated Rhythms: - Asynchronous @ 75 BPM
- Demand with frequent sinus beat
- Demand with occasional sinus beat
- A-V sequential
- Non-capture
- Non-function
• Pacer Pulse Amplitude: 2, 4, 6, 8, 10, 12, 14, 16, 18, 20, 50,
100, 200, 500 and 700 mV
• Pacer Pulse Width: 0.1, 0.2, 0.5, 1.0, and 2.0 msec
• Pacer Pulse Polarity: Positive or negative
Invasive Blood Pressure:
• Channels: 2 independent channels
• Accuracy: ± (1% s 1 mmHg)
• Static Pressures: -10 to 400 mmHg; 1 mmHg increments
15 customizable presets
• Transducer Sensitivity: 5 µV/V/mmHg or 40 µV/V/mmHg
• I/O Impedance: 300 ohms ± 10%
• Excitation Voltage: 2 to 16 Vp; 0 to 5 kHz
• Dynamic Simulations: - Arterial 120/80, 90/40, and 160/110
- Radial Artery 120/80
- Left Ventricle 120/0
- Right Ventricle 25/0
- Pulmonary Artery 25/10
- Pulmonary Wedge 10/2
- Right Atrium (CVP) 120/0
- Left Atrium 14/4
• Swan-Ganz Simulations:
(Channel 1 only)
Auto or Manual
• Blood Pressure Artifact: OFF
5% for Arterial, Radial and L. Ventricle, or 5 mmHg
for others
10% for Arterial, Radial and L. Ventricle, or 10
mmHg for others
Specifications/Chapter 1
#
Page 7
Содержание vPad-A1
Страница 1: ...Operating Manual MN 103c vPad A1 All in One Patient Simulator and Performance Tester ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 14: ...vPad A1 OPERATORS MANUAL Definitions Page x ...
Страница 28: ...vPad A1 OPERATORS MANUAL Specifications Chapter 1 Page 14 ...
Страница 32: ...vPad A1 OPERATORS MANUAL Overview Chapter 2 Page 18 ...
Страница 102: ...vPad A1 OPERATORS MANUAL Automation and Test Records Chapter 5 Page 88 ...
Страница 104: ...vPad A1 OPERATORS MANUAL Maintenance Chapter 6 Page 90 ...
Страница 107: ...vPad A1 OPERATORS MANUAL Appendix A Page 93 ...
Страница 108: ......