vPad-A1 OPERATORS MANUAL
The
button can abort the
Leakage Rate
test at any time. The test can then be restarted
with the
button on the setup page.
At the beginning of an
Auto Mode
test, the A1 will begin pumping air into the pneumatic system
until pressure reaches
Start Pressure
. The
Test Time
countdown will automatically start after the
target pressure is reached.
In
Manual Mode
, any incoming pressure must be from an external source (i.e., from the NIBP
monitor itself, or alternatively from a hand pump). When pressure reaches the desired level, then
users can start the
Test Time
countdown by pressing the
button on the
Result
screen.
For both
Modes
, the
button is initially disabled; however, users can assign
if, for
example, it becomes clear that a system leak is too large to reach
Start Pressure
. Otherwise, the
PASS result will be available at the conclusion of the test.
When
Test Time
expires, the A1 measures the pressure one last time, then
vents the pneumatic system. The test will then calculate and display the
Leakage Rate
in mmHg per minute.
Figure 93
Manual Operation/Chapter 4
#
Page 64
Содержание vPad-A1
Страница 1: ...Operating Manual MN 103c vPad A1 All in One Patient Simulator and Performance Tester ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 14: ...vPad A1 OPERATORS MANUAL Definitions Page x ...
Страница 28: ...vPad A1 OPERATORS MANUAL Specifications Chapter 1 Page 14 ...
Страница 32: ...vPad A1 OPERATORS MANUAL Overview Chapter 2 Page 18 ...
Страница 102: ...vPad A1 OPERATORS MANUAL Automation and Test Records Chapter 5 Page 88 ...
Страница 104: ...vPad A1 OPERATORS MANUAL Maintenance Chapter 6 Page 90 ...
Страница 107: ...vPad A1 OPERATORS MANUAL Appendix A Page 93 ...
Страница 108: ......