vPad-A1 OPERATORS MANUAL
1.2 Multi-Parameter Patient Simulation
Multi-parameter patient simulation is provided by the
vPad-PS
module of vPad-A1.
ECG General:
• Simulation Type: Full 12-Lead ECG with independent outputs for each
signal lead, referenced to RL
• Output Impedance: 500, 1000, 1500, and 2000 ohms to RL
• High Level Output: Lead II amplitude x 500 (0.5V per Lead II millivolt)
• Amplitude Accuracy: ± 2% (Lead II, 2 Hz Square Wave)
• QRS width: 80 msec (Adult) or 40 msec (Neonatal)
• Artifact: 50 Hz, 60 Hz, Muscle, Baseline (wander), Respiration
C
Rate Accuracy Better than 0.1%
Normal Sinus Rhythm:
• Rates: 10 to 360 BPM
presets: 20, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220,
240, 260, 280, and 300 BPM (default/user re-defineable)
• Rate Accuracy: Better than 0.1%
• Amplitudes (Lead II): 0.05 to 0.50 mV in 0.05 mV steps, 0.5 to 5.5 mV in 0.25
mV steps
• Axis Deviation: Normal (intermediate), horizontal, and vertical
Modifies ECG baseline during arrhythmias
Specifications/Chapter 1
#
Page 3
Содержание vPad-A1
Страница 1: ...Operating Manual MN 103c vPad A1 All in One Patient Simulator and Performance Tester ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 14: ...vPad A1 OPERATORS MANUAL Definitions Page x ...
Страница 28: ...vPad A1 OPERATORS MANUAL Specifications Chapter 1 Page 14 ...
Страница 32: ...vPad A1 OPERATORS MANUAL Overview Chapter 2 Page 18 ...
Страница 102: ...vPad A1 OPERATORS MANUAL Automation and Test Records Chapter 5 Page 88 ...
Страница 104: ...vPad A1 OPERATORS MANUAL Maintenance Chapter 6 Page 90 ...
Страница 107: ...vPad A1 OPERATORS MANUAL Appendix A Page 93 ...
Страница 108: ......