vPad-A1 OPERATORS MANUAL
When ready to begin testing a patient monitor, power ON the system by depressing the pushbutton switch
located at upper right on the
Base
. The power LED at top left on the
Base
should light up and display
green
.
Switch ON the
Tablet
by pressing and holding the Power / Screen Lock button for about 5 seconds. The
Tablet
should boot up and eventually show the Home screen within 20 seconds. Launch the
vPad-A1 App
to set up and operate vPad-A1; the patient simulator is controlled wirelessly, via Bluetooth.
Note:
Once the patient simulation has been set up with the
App
, it is good practice to turn the
Tablet
display OFF temporarily. Quickly pressing and releasing the Power / Screen Lock button on the
Tablet
will put its screen into a ‘sleep’ mode. The ‘sleep’ state is not the same as OFF, but it will
reduce power consumption and extend battery life of the
Tablet
. Momentarily press the same
button again to ‘wake’ the user interface.
Figure 8
Figure 9
Figure 10
Setup/Chapter 3
#
Page 24
Содержание vPad-A1
Страница 1: ...Operating Manual MN 103c vPad A1 All in One Patient Simulator and Performance Tester ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 14: ...vPad A1 OPERATORS MANUAL Definitions Page x ...
Страница 28: ...vPad A1 OPERATORS MANUAL Specifications Chapter 1 Page 14 ...
Страница 32: ...vPad A1 OPERATORS MANUAL Overview Chapter 2 Page 18 ...
Страница 102: ...vPad A1 OPERATORS MANUAL Automation and Test Records Chapter 5 Page 88 ...
Страница 104: ...vPad A1 OPERATORS MANUAL Maintenance Chapter 6 Page 90 ...
Страница 107: ...vPad A1 OPERATORS MANUAL Appendix A Page 93 ...
Страница 108: ......