vPad-A1 OPERATORS MANUAL
The
buttons
in
Figure 105
change the selected point's
pressure, and
buttons change its amplitude.
The
AMAP
button displays the current maximum pulse height, which
defines the largest simulated pulse that vPad-BP will produce during
an
NIBP Simulation
. If the device requires smaller pulses (e.g. the
test cuff is smaller than the typical adult cuff), then the
AMAP
controls can be used to adjust the maximum pulse height as needed.
The
Shift
controls move the entire
Envelope
horizontally along the pressure axis. Pressing a
or
button moves all inflection points by the same pressure shift. The default increment is 10
mmHg of pressure, though users can enter any custom number from 0 to 50 in the input box.
Changes to the
Shift
increment will only be applied after pressing the "Done" button on the
keyboard.
Pressing the
button in
Figure 103
dismisses the tool and returns to NIBP simulation
settings after warning users of unsaved changes.
After users have verified via the
button that the displayed
Envelope
produces the expected readings on the target device, the
button will launch a “Save Envelope” form.
The form is pre-populated with information extracted from the
Envelope
template originally selected for editing. The
Manufacturer
and
Model
fields should contain the device make and model,
respectively, both of which can be edited if necessary.
Pressing the
Mode
button toggles between "Adult" and "Neonatal"
mode. This should be determined by the targeted monitor type and
settings used during
Envelope
customization.
The
Targets
fields contain three NIBP measurements (SYS-DIA-
MAP) that will later inform users on what blood pressure readings
they should expect from a properly-functioning device of the same
make and model. The automatically generated
Envelope
name,
which will be listed in the
Simulation
dropdown menu, is derived
from the entered
Targets
.
Pressing
in
Figure 106
saves all displayed pressure parameters to an
Envelope
file. If
none of the fields in the form were edited, then the originally chosen template will be overwritten.
Figure 105
Figure 106
Manual Operation/Chapter 4
#
Page 71
Содержание vPad-A1
Страница 1: ...Operating Manual MN 103c vPad A1 All in One Patient Simulator and Performance Tester ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 14: ...vPad A1 OPERATORS MANUAL Definitions Page x ...
Страница 28: ...vPad A1 OPERATORS MANUAL Specifications Chapter 1 Page 14 ...
Страница 32: ...vPad A1 OPERATORS MANUAL Overview Chapter 2 Page 18 ...
Страница 102: ...vPad A1 OPERATORS MANUAL Automation and Test Records Chapter 5 Page 88 ...
Страница 104: ...vPad A1 OPERATORS MANUAL Maintenance Chapter 6 Page 90 ...
Страница 107: ...vPad A1 OPERATORS MANUAL Appendix A Page 93 ...
Страница 108: ......