vPad-A1 OPERATORS MANUAL
It can be set from 0.05 to 1.0 ohms in 0.05 ohm increments, and from 1.0 to 5.0 ohms in 0.25 ohm
increments.
4.3.2.4 Lead Selection
The
Leads
parameter selects the ECG leads across which the simulated
Impedance Variation
is
applied. Choices are between the RA and LL leads, or alternatively between the RA and LA leads.
4.3.2.5 Inspiration/Expiration Ratio
The
Inspiration/Expiration Ratio
(
I:E
) defines a relationship between inspiration (inhalation) time and
expiration (exhalation) time, which is an indicator of patient respiratory effort. It can be set to one
of the following options:
1:1, 1:2, 1:3, 1:4, 1:5, or Ventilated 1:1
4.3.2.6 Apnea
Apnea
is the temporary cessation of breathing, which may occur while sleeping or under anaesthesia.
This setting imitates the condition by suspending
Respiration
simulations for the selected duration.
Available settings are:
Off, Continuous, 12 seconds, 22 seconds, or 32 seconds
Note
: ‘Continuous’ means breathing is stopped until another
Apnea
setting is chosen.
Manual Operation/Chapter 4
#
Page 42
Содержание vPad-A1
Страница 1: ...Operating Manual MN 103c vPad A1 All in One Patient Simulator and Performance Tester ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 14: ...vPad A1 OPERATORS MANUAL Definitions Page x ...
Страница 28: ...vPad A1 OPERATORS MANUAL Specifications Chapter 1 Page 14 ...
Страница 32: ...vPad A1 OPERATORS MANUAL Overview Chapter 2 Page 18 ...
Страница 102: ...vPad A1 OPERATORS MANUAL Automation and Test Records Chapter 5 Page 88 ...
Страница 104: ...vPad A1 OPERATORS MANUAL Maintenance Chapter 6 Page 90 ...
Страница 107: ...vPad A1 OPERATORS MANUAL Appendix A Page 93 ...
Страница 108: ......