vPad-A1 OPERATORS MANUAL
4.2.10 ST-Segment Elevation
ST-Segment Elevation
simulates regular NSR waves with ST-segments that are offset from the baseline. Of
the common ECG parameters, only
Amplitude
and
Artifact
are available for this test. The
Rate
is fixed at
60 BPM.
The percentage of ST-segment offset, relative to the ECG
Amplitude
, is adjustable by changing the
ST Level
setting to a value between -150% (below baseline) and +150% (above baseline).
4.2.11 Tall T-Wave
The
Tall T-Wave
simulation generates large T-waves accompanying standard QRS waves.
Per ANSI/AAMI EC-13, the QRS wave has a width of 100 ms and a QT interval of 350 ms. The T-wave
has a sinusoidal shape, 180 ms wide, and
T Height
can be adjusted in amplitude from 0 to 150 % of the QRS
waveform amplitude.
The QRS waveform amplitude is adjustable by changing the ECG
Amplitude
parameter. The
Rate
for
Tall
T-Wave
s is fixed at 80 BPM.
Manual Operation/Chapter 4
#
Page 40