ES601-US OPERATING MANUAL
Manual Testing/Chapter 5
#
Page 71
Figure 39
- Leakage test paused during sequence.
The type of measurement being performed and the AC supply conditions are shown on the display, above
the leakage readings. RMS and DC limits corresponding to the test conditions are also shown on the right,
above the readings. Each RMS and DC value is respectively compared to its limit and is either "PASSED"
or "FAILED" automatically by ES601-US. The
PAUSE
button may be pressed at any time to pause the
sequence and freeze the most recent measurement on the display (see
Figure 37
).
Individual leakage readings are rapidly made by ES601-US, each reading usually requiring less than one
second. If it is expected you may want to pause individual values during a test in order review them, you
should set the "PAUSE AFTER..." setting on the Sequence Control menu (
Figure 34
) to at least 2 seconds,
so the time between readings is increased to allow for user reaction time. "PAUSE AFTER..." can be set
from 1 to 5 seconds in steps of one second, or the pause can be completely disabled, allowing you to trade
off operation of the
PAUSE
button against total execution time of the leakage test.
Содержание ES601-US
Страница 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Страница 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Страница 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Страница 60: ...ES601 US OPERATING MANUAL Connecting To ES601 Chapter 4 Page 40 Figure 17 Connecting to a line powered DUT ...
Страница 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Страница 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Страница 149: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 129 Figure 95 Checklist in progress A Checkitem ...
Страница 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Страница 208: ...ES601 US OPERATING MANUAL ES601pc US Software Chapter 8 Page 188 Figure 137 Record viewing options ...
Страница 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Страница 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Страница 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Страница 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Страница 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...