ES601-US OPERATING MANUAL
Manual Testing/Chapter 5
#
Page 84
Figure 48
: Example of patient leakage results (1 page per Group).
For each supplementary test condition enabled by the user in the test setup, ES601-US measures all
combinations of the supplementary condition with the AC supply conditions specified in the safety
standard. ES601-US then records the AC and DC leakage results for the supplementary condition that
produced the highest leakage. The supplementary condition producing the worst case is also recorded and
appears in parentheses in the Test Record report.
If the user chose to set up the test so certain conditions were disabled, then AC and DC measurement
results for the disabled tests appear in the Record as "-----". For conditions pertaining to the AC supply
(OPEN N., OPEN G.), "-----" results indicate the user may have inadvertently forgotten to enable the test.
This may prompt the user to repeat the test with the correct conditions.
For all safety standards except IEC61010, patient leakage results will follow enclosure leakage results if the
user has enabled testing of patient leads. One page is provided for each functional group of patient leads, as
configured by the user in the test setup (see example below).
For the IEC60601 safety standard, the patient leads of a "Type CF" group are tested individually in turn. On
the patient leakage page, in the "AP" column ES601-US shows the patient connection that produced the
highest leakage for the given test condition. For "Type B" and "Type BF" groups, or for testing of any
"Type" done under the IEC62353 safety standard, the leads are tested collectively and so "
ALL
" will appear
in the "AP" column of the patient leakage results.
Содержание ES601-US
Страница 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Страница 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Страница 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Страница 60: ...ES601 US OPERATING MANUAL Connecting To ES601 Chapter 4 Page 40 Figure 17 Connecting to a line powered DUT ...
Страница 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Страница 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Страница 149: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 129 Figure 95 Checklist in progress A Checkitem ...
Страница 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Страница 208: ...ES601 US OPERATING MANUAL ES601pc US Software Chapter 8 Page 188 Figure 137 Record viewing options ...
Страница 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Страница 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Страница 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Страница 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Страница 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...