ES601-US OPERATING MANUAL
Manual Testing/Chapter 5
#
Page 68
Figure 36
- Sequence Controls for manual leakage test.
The buttons provided on the Sequence Control menu are:
PAUSE AFTER
MEASUREMENT:
After each measurement of the test, ES601-US will pause for the indicated time
before proceeding with the next leakage measurement. The delay provided by
this control allows the user to more easily interrupt a sequence in progress, so a
particular leakage measurement of interest can be repeated as desired. This
control can be set to "NO" to disable the pause.
POWER UP DELAY:
During the test, ES601-US will wait for the indicated time when the AC supply
is applied to the DUT, before proceeding with the sequence. This delay allows
the DUT to start up and stabilize before subsequent leakage measurements are
taken. The available options are 2 to 9 seconds in one second steps. Equipment
requiring more than 9 seconds to start up are better accommodated through the
"STOP AFTER..." control, described following.
STOP AFTER
SUPPLY ON:
If set to "YES", this control will override the power-up delay setting described
prior and will cause the test to
fully stop
immediately after the AC supply is
connected to the DUT by ES601-US. ES601-US will prompt the user to
activate the DUT and will not proceed until the user acknowledges by pressing
the
CONTINUE
button. This feature accommodates equipment with a
momentary ON/OFF power switch, which cannot be switched ON until a full
AC supply is provided to the equipment.
Содержание ES601-US
Страница 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Страница 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Страница 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Страница 60: ...ES601 US OPERATING MANUAL Connecting To ES601 Chapter 4 Page 40 Figure 17 Connecting to a line powered DUT ...
Страница 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Страница 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Страница 149: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 129 Figure 95 Checklist in progress A Checkitem ...
Страница 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Страница 208: ...ES601 US OPERATING MANUAL ES601pc US Software Chapter 8 Page 188 Figure 137 Record viewing options ...
Страница 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Страница 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Страница 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Страница 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Страница 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...