ES601-US OPERATING MANUAL
Manual Testing/Chapter 5
#
Page 70
Figure 38
- Leakage current test in progress.
WARNING:
If testing of
Type BF
or
Type CF
patient leads has been enabled, HIGH VOLTAGE (isolation test
voltage) will be periodically output from the AP jacks during the test.
The isolation test voltage is isolated from the AC mains and is internally current-limited to 8 milliamperes
maximum at 264 VAC output. Unterminated leads attached to the AP jacks therefore present a minor,
non-lethal shock hazard to the careless user.
Always take care to make proper electrical connections to the DUT before starting a leakage test.
Disconnect all unused leads or cables from the AP jacks on ES601-US.
Once the leakage test has been set up and the connections are made, the test is started by pressing the
CONTINUE
button on the CONNECTIONS menu. The required measurements are performed
automatically in a sequence. AC RMS and DC currents are measured simultaneously, and the display is
updated with new data approximately two times every second. If the "PAUSE AFTER..." Sequence Control
is not set to "NO", ES601-US will delay an additional 1 to 5 seconds between consecutive leakage
measurements.
Содержание ES601-US
Страница 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Страница 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Страница 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Страница 60: ...ES601 US OPERATING MANUAL Connecting To ES601 Chapter 4 Page 40 Figure 17 Connecting to a line powered DUT ...
Страница 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Страница 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Страница 149: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 129 Figure 95 Checklist in progress A Checkitem ...
Страница 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Страница 208: ...ES601 US OPERATING MANUAL ES601pc US Software Chapter 8 Page 188 Figure 137 Record viewing options ...
Страница 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Страница 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Страница 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Страница 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Страница 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...