ES601-US OPERATING MANUAL
Overview/Chapter 2
#
Page 16
Patient Leakage (record for each Group defined by the user):
Normal Polarity
Normal Condition
(2)
Single-Fault Condition, Isolation Test
(NP:NORM.)
(NP:ISO. TEST)
Reverse Polarity
Normal Condition
(2)
Single-Fault Condition, Isolation
Test(s)
(RP:NORM.)
(RP:ISO. TEST)
(2) VDE-0751 only: normal condition (patient lead(s) to ground) is measured if user enables "per IEC60601"
method in the test setup.
2.1.11 Battery-Power Protocols
If the DUT is an internally-powered device, and cannot be operated from or connected to the AC line, then
a subset of the above tests is applicable. If the DUT is defined as battery-powered by the user, ES601-US
will test only patient lead leakage and lead-to-lead leakage of the DUT, and measure insulation resistance of
the patient leads to ground.
To accommodate the unique needs of battery-powered devices, ES601-US incorporates special protocols
designated "AAMI-BAT" (AAMI-ES1 standard for self-powered device), "IEC601-BAT" (IEC60601
standard for self-powered device), "VDE-BAT" (VDE-0751 standard for self-powered device) and
"IEC353-BAT" (IEC62353 standard for self-powered device).
2.1.12 Safety AutoSequences
ES601-US can be programmed to automatically conduct the safety measurements described above.
Automated safety measurements are always performed by ES601-US in a predefined order. The settings of
an ES601-US Safety AutoSequence determine which measurements within the sequence are enabled, but do
not affect the order in which the measurements are conducted.
ES601-US provides memory for storage of 24 user-programmable Safety AutoSequences. These sequences
can be programmed and saved using the operator interface of ES601-US. Alternatively, the sequences can
be programmed and loaded into ES601-US memory by means of the Safety AutoSequence module of the
ES601pc-US software package (see Chapter 8). ES601pc-US software is included on CD ROM with ES601-
US or may be downloaded or upgraded from Datrend’s website -
www.datrend.com
.
A Safety AutoSequence can include measurements of line voltage, load current, ground resistance, insulation
resistance, and leakages (external case, internal case, patient leads, lead-to-lead, etc.). For equipment having
patient leads, each Safety Autosequence can be set up to organize the parts into
Functional Groups
of defined
Protection Class
, with names assignable to each Group as well as to each lead within each Group. Up to ten
Содержание ES601-US
Страница 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Страница 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Страница 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Страница 60: ...ES601 US OPERATING MANUAL Connecting To ES601 Chapter 4 Page 40 Figure 17 Connecting to a line powered DUT ...
Страница 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Страница 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Страница 149: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 129 Figure 95 Checklist in progress A Checkitem ...
Страница 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Страница 208: ...ES601 US OPERATING MANUAL ES601pc US Software Chapter 8 Page 188 Figure 137 Record viewing options ...
Страница 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Страница 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Страница 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Страница 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Страница 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...