ES601-US OPERATING MANUAL
Manual Testing/Chapter 5
#
Page 67
GROUP
STATUS:
Selects "ENABLED" or "DISABLED". Patient leads of disabled groups are excluded
from patient leakage and lead-to-lead leakage measurements. Group 1 cannot be
disabled.
GROUP
TYPE:
Selects the type of group: "B NON-ISOLATED", "BF ISOLATED" (default), or "CF
ISOLATED". Refer to Chapter 2 to review the meaning of "Type B", "Type BF", and
"Type CF" protection classes of the safety standard. The setting of GROUP TYPE
affects the limits applied in patient leakage tests.
NAME OF
GROUP:
Pressing this button causes the QWERTY touch keypad to be displayed. The user may
enter up to seven letters and/or digits as a description of the lead group.
NUMBER OF
LEADS:
Determines the number of leads in the group being displayed. Leads are incorporated
into the group consecutively, starting at jack AP1 on the front panel. The simplest
permitted setup is one patient lead (AP1) in one enabled group (Group 1).
BACK:
Returns to the PATIENT LEADS menu.
RESET ALL:
Resets all settings, leaving Group 1 enabled as Type BF with one patient lead (AP1).
If the DUT is a defibrillator and has been specified by the manufacturer as meeting the requirements of the
IEC60601-2-4 standard ("
Particular requirements for the safety of cardiac defibrillators
"), group settings can be
programmed via the GROUPS editor menu to enable special leakage tests which are required by that
standard. For testing a defibrillator per IEC60601-2-4:
• Assign ECG leads to one group, and set the GROUP TYPE to "CF ISOLATED"
• Assign the defibrillator paddles to a second group, and set NUMBER OF LEADS equal to two.
Set the GROUP TYPE to "BF ISOLATED" or "CF ISOLATED" according to manufacturer
specifications. Finally, enter "PADDLES", "PIASTRE" or "PALETTE" for the NAME of the
group.
The above special settings will enable ES601-US to perform leakage testing of the paddles per IEC60601-2-
4, with modified acceptance limits according to that standard.
After configuring the leakage test conditions and, if necessary, specifying the DUT patient leads, the next
step displays the Sequence Controls for manual leakage tests, as recalled from System Settings Menu #2 (see
Section 3.3). You may change the displayed settings to suit the upcoming test. After the test is completed,
the Sequence Controls will revert to their System Setting defaults.
Содержание ES601-US
Страница 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Страница 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Страница 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Страница 60: ...ES601 US OPERATING MANUAL Connecting To ES601 Chapter 4 Page 40 Figure 17 Connecting to a line powered DUT ...
Страница 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Страница 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Страница 149: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 129 Figure 95 Checklist in progress A Checkitem ...
Страница 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Страница 208: ...ES601 US OPERATING MANUAL ES601pc US Software Chapter 8 Page 188 Figure 137 Record viewing options ...
Страница 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Страница 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Страница 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Страница 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Страница 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...