ES601-US OPERATING MANUAL
Computer Control/Chapter 7
#
Page 139
7.4.4 Format of Measurement Data
Once initiated by computer control as previously described, ES601-US will transmit one or two lines of
ASCII text for each measurement completed during the test sequence. Two lines of data are sent for leakage
measurements (AC result, followed by DC result); for all other measurements ES601-US sends one line of
data. Each line of data is terminated with a carriage return
\r
(0x0D). Measurement data is separated into
fields by the comma character
,
(0x2C). This type of format is referred to as "comma-separated variables"
(CSV) and files containing such data are directly compatible with Microsoft® Excel.
The first field of the CSV data is a 3-character mnemonic which specifies the type of measurement
performed. Depending on the type of measurement, the acceptance limit specified in the standard may be
included in the data, and if so, the last field before the terminating carriage return
\r
will be either a
P
(pass) or
F
(fail) character as appropriate.
For each measurement performed by ES601-US, the format of the return data is listed in the following
table.
Safety standard
STD,[standard]\r
Line voltages
ACV,<V
L1-L2
>,<V
L1-GND
>,<V
L2-GND
>,Vrms\r
Equipment
load current
CUR,<current>,Arms\r
Insulation,
L1+L2 to ground
ILG,<value>,<limit>,Megohm,[-|P|F]\r
Insulation,
AP to ground
IAP,<value>,<limit>,Megohm,[-|P|F]\r
Ground
PRE,[ 1A|25A],[LC|PP],<value>,<limit>,ohm,[P|F]\r
Leakage, Encl.,
IEC60601/AAMI
LEN,O??/?P/L2?/E?/AP?/X1?/X2?,<value>,<limit>,[uArms|uAdc],[P/F]\r
Leakage, Earth,
IEC60601/AAMI
LEA,O??/?P/L2?/E?/AP?/X1?/X2?,<value>,<limit>,[uArms|uAdc],[P|F]\r
Leakage, Patient,
IEC60601/AAMI
LPA,O??/?P/L2?/E?/AP?/X1?/X2?,GPx,APyy,<value>,<limit>,[uArms|uAdc],[P|F]\r
Leakage, Auxiliary,
IEC60601
LAX,O??/?P/L2?/E?,APyy,<value>,<limit>,[uArms|uAdc],[P|F]\r
Leakage, Interlead,
AAMI
LLD,O??/?P/L2?/E?,APyy,<value>,<limit>,[uArms|uAdc],[P|F]\r
Содержание ES601-US
Страница 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page iv ...
Страница 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Страница 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Страница 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Страница 60: ...ES601 US OPERATING MANUAL Connecting To ES601 Chapter 4 Page 40 Figure 17 Connecting to a line powered DUT ...
Страница 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Страница 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Страница 149: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 129 Figure 95 Checklist in progress A Checkitem ...
Страница 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Страница 208: ...ES601 US OPERATING MANUAL ES601pc US Software Chapter 8 Page 188 Figure 137 Record viewing options ...
Страница 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Страница 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Страница 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Страница 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Страница 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...