7-4
Technical Data
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF IEC
61000-4-3
3 Vrms 150kHz to
80MHz
3 V/m 80MHz to
2.5GHz
3 V
3 V/m
Portable and mobile
RF communications
equipment should be used
no closer to any part of
the Angel
®
Concentrated
Platelet Rich Plasma
(cPRP) System,
including cables, than the
recommended separation
distance calculated from
the equation applicable to
the frequency of
the transmitter.
Recommended
separation distance
d=1.17√P
d=1.17√P
80MHz to 800MHz
d=2.33√P
800MHz to 2.5GHz
where
P
is the maximum
output power rating
of the transmitter in
Watts (W) according
to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
survey
i
, should be less
than the compliance level
in each frequency range
ii
.
Interference may occur
in the vicinity of
equipment marked with
the following symbol:
Guidance and manufacturer’s declaration – electromagnetic immunity
The Angel
®
Concentrated Platelet Rich Plasma (cPRP) System is intended for
use in the electromagnetic environment specified below. The customer or the
user of the Angel
®
Concentrated Platelet Rich Plasma (cPRP) System should as-
sure that it is used in such an environment.
Note: At 80MHz and 800MHz, the higher frequency range applies.