Angel
®
cPRP System
Operator’s Manual
v
Before You Get Started
12. The Angel
®
cPRP Processing Set is intended for single patient use. Each set
can be used on the same patient for up to three processing cycles. Do not
resterilize any part of this Processing Set.
13. Failure to properly load the Angel
®
cPRP Processing Set per the enclosed
instructions may affect the performance of the system.
14. Luer lock syringes should be used with the Angel
®
cPRP Processing Set.
15. Pressing the Stop button during separation may reduce processing efficiency.
16. The physician ordering the collection of PRP shall use discretion when any of
the following conditions exist:
• sepsis
• preoperative hematocrit less than 30%
• preoperative platelet count less than 195,000 per ul
• hemodynamically unstable
• prolonged clotting times
• recent use of anti-platelet drugs
• inability to maintain stable oncotic pressure
17. Attach the power cord only to a power receptacle which is properly grounded.
18. Replace the mains fuses only with fuses of the same type and rating.
19. There are no user-serviceable parts inside this device. To avoid the risk of
electrical shock, do not remove the cover. Refer all servicing to qualified
service personnel.
20. Take care not to misplace the screws and springs when disassembling the
centrifuge adaptor plate. If springs or screws are misplaced, you may order
replacements by contacting Cytomedix, Inc.
21. Federal law (USA) restricts this device to sale by or on the order of a physician.
22. The user of the Angel
®
Concentrated Platelet Rich Plasma (cPRP) System
is responsible for the monitoring of the system performance when using
custom protocols.
23. The Angel
®
Concentrated Platelet Rich Plasma (cPRP) System has been
tested and verified to meet applicable Electromagnetic Compatibly (EMC)
and Electrical Safety standards when used in conjunction with the specified
Data Integrity Accessories. (Refer to Section 6.)
24. The Angel
®
Concentrated Platelet Rich Plasma (cPRP) System has the
ability to save data using both the USB and Ethernet Network connections.
When using either of these connections, it is recommended that the
applicable Data Integrity Accessory be used. (Refer to Section 6.)