Angel
®
cPRP System
Operator’s Manual
5-2
Troubleshooting
Name
Icon
Description
Response
Insufficient fill
The insufficient fill
alarm is triggered if air
is detected before the
desired amount of whole
blood or mixture of
blood and bone marrow
has been pumped into
the variable volume
separation chamber.
Along with the alarm,
two buttons are
displayed on the Run
Screen: the “Process”
button and the
“Continue Filling” button.
If more than 40 ml of
blood or mixture of
blood and bone marrow
has been pumped into
the variable volume
separation chamber,
you may process it
by touching the
“Process” button.
Or, if available whole
blood/bone marrow is
available in the whole
blood compartment and
you wish to continue
filling to the desired
volume, check to make
sure there are no breaks
or occlusions in the
line and then press the
“Continue Filling” button
to continue the cycle.
Lid latch
If the latch is detected
to be unlatched after the
Start button is pressed,
the centrifuge is stopped
and the Lid latch alarm
is sounded.
This alarm indicates a
hardware failure. If it
does not resolve itself,
contact the Cytomedix,
Inc. Customer Service
Response Center.
Lid open
yellow
red
If the lid is open before
the Start button is
pressed, the yellow icon
appears, prompting you
to close the lid properly.
If the lid opens during
the cycle, the red
icon appears (and the
centrifuge stops until the
problem is corrected).
Close the lid cover.
Ensure tubing is not
caught between lid and
body of the machine.