Angel
®
cPRP System
Operator’s Manual
3-13
Processing
Processing
3-14
Angel Operator’s Manual
8. Menu Button
Touching the Menu button will bring up the Menu Screen. Various Angel
settings can be adjusted from the Menu Screen. Touching the Close
button in the Menu Screen will close the Menu Screen and display the previously
displayed screen. Refer to the section “Menu Screen” on page 3-19.
9. Info Button
Touching the Info button from any screen may bring up an Info Screen
containing contextual help for that screen. Touching the Close button from
the Info Screen will return you to the previously displayed screen. Refer to the
section “Info Screen” on page 3-22.
Run Screen
The Run Screen tracks the progress of the blood separation process by
dynamically displaying (in ml) the amount of RBC, PPP, and PRP collected
throughout the processing cycle (see Figure 3-11).
A tally of the total whole blood processed and the RBC, PPP, and PRP collected
during a case may be viewed at any time by touching the Menu button, and then
touching the "Tally" tab from the Menu Screen (see Figure 3-15).
Figure 3-11 Run Screen
Pause/Resume button
Processing may be paused at any time from the Run Screen by
touching the Pause/Resume button (the yellow bars). When the
system is paused, the word “Paused” appears below the
processing time.
2
1
1 Pause/Resume button
2 Modify Volume button
Run Screen
The Run Screen tracks the progress of the blood separation process by
dynamically displaying (in ml) the amount of RBC, PPP, and PRP collected
throughout the processing cycle (see Figure 3-11).
A tally of the total whole blood/bone marrow processed and the RBC, PPP, and
PRP collected during a case may be viewed at any time by touching the Menu
button, and then touching the “Tally” tab from the Menu Screen (see Figure 3-15).
1. Pause/Resume Button
2. Modify Volume Button
Figure 3-13 Run Screen
Pause/Resume Button
Processing may be paused at any time from the Run Screen
by touching the Pause/Resume button (the yellow bars). When
the system is paused, the word “Paused” appears below the
processing time.
A paused processing cycle may be resumed by touching the Pause/Resume
button a second time (the green arrow).
Modify Volume Button
The volume of whole blood or mixture of blood and bone marrow which
has been selected to process is displayed within the Modify Volume
button (ml). Touching the Modify Volume button allows the operator to
change this volume during the fill stage of processing.
Touching the Modify Volume button pauses the system and displays a screen
similar to the Start Screen where the operator may use the Up Arrow button and
Down Arrow button to increase or decrease, respectively, the volume of whole
blood or mixture of blood and bone marrow to process (refer to “Start Screen” on
page 3-11). A volume less than that already within the variable volume separation
chamber cannot be selected.
Once the new volume has been selected, touching the green Resume button will
continue the processing cycle.