Angel
®
cPRP System
Operator’s Manual
5-6
Troubleshooting
Text
Description
Response
No software updates found!
No software updates were found. See “Troubleshooting the Software
Update Process” on page 5-8 for
troubleshooting steps.
No USB storage device found!
The Angel
®
cPRP System does
not detect any connected USB
storage device.
If you wish to output tally data
or a log file to a USB storage
device, refer to the section
“Troubleshooting the Save
Process” on page 5-8 for
troubleshooting steps.
Output failed!
Output failed while attempting to
save tally data.
Refer to the section
“Troubleshooting the Save
Process” on page 5-8 for
troubleshooting steps.
Save failed!
Output failed while attempting to
save case log file. Or a custom
protocol could not be saved.
Refer to the section
“Troubleshooting the Save
Process” on page 5-8 for
troubleshooting steps.
Past case data lost!
Some or all of the data for past
or current cases has been lost.
No response necessary. Past
case data may not be recovered.
Powerfail protocol not found:
xxxxxx
A powerfail has occurred, and
the given protocol cannot be
found.
The Angel
®
cPRP System will
automatically select the default
protocol for use. If a different
protocol is desired, select it from
the Protocol pull-down button.
The lost protocol may have to be
recreated before it can be used.
Power was lost before end of
case.
The Angel
®
cPRP System was
shut down or lost power before
the end of a processing case.
If power was lost in the middle
of a cycle, then the set must be
emptied. A new cycle or case
may then be started.
Redo failed! Undo/redo history
cleared.
An attempt to redo a change
while creating a protocol has
failed.
The undo/redo history has been
cleared. A redo command is not
possible at this time.
Rename failed!
An attempt to rename a protocol
has failed.
Try renaming the protocol again.
Settings lost and were set to
default!
The Angel
®
cPRP System’s
settings were lost, and the
default settings are being used.
If settings other than the default
are desired, make changes via
the "Settings" tab of the Menu
Screen (see “Settings Tab” on
page 3-21).
Software update is not valid!
The new software downloaded
from the selected source is
corrupt.
Try downloading the software
again. If the problem persists,
download software from a
different source.
The software did not start
successfully. Turn power off
and on. If the problem persists,
download software.
An error occurred while the
Angel
®
cPRP System was
powering on.
Turn power off and then on. If
the problem persists, contact
the Cytomedix, Inc. Customer
Service Response Center.
This version of software is
already installed on this machine.
The operator has attempted to
install the same software which is
already installed on the machine.
Choose a different version of
software to install.