6-10
Routine Care
1. Using a flat screwdriver, pry open the covering on the power switch.
2. Once you have opened the cover, the two fuse holders are visible.
Note the orientation of the downward arrows on the fuse holder. Pull to
remove each holder from the machine.
3. Once you remove the fuse holders, replace the fuses themselves with fuses
of identical specification (5X20mm size, 5A 250VAC fast blow).
4. Reverse the removal steps in order to re-install.
Verify that the stator arm is not loose on its pivot pin. It should only rotate up
and down with hand pressure and have no side-to-side motion. If the arm is
loose, lift it to the raised braking position and locate the cap screw near the
base of the stator arm. Using a 9/64 inch hex wrench, tighten the cap screw
against the pivot pin.
1. Power Switch cover conceals the two fuses
Figure 6-10 Fuse Replacement: Pry Open Cover
Figure 6-11 Fuse Replacement: Slide Out Holders