25
0473
90001
Rev
07.26.13
Mounting System Stability
Inspect Beam Collimator Device
Serial Number Label
Physical Damage
Inspect HV (X-Ray) Power Supply
Serial Number Label
Visual Exposure Indicator
Audible Exposure Signal
Check Platform Level
Level with system feet, starting with back 2
Check Laser Alignment lights
Vertical Cross hair to vertical center of panel
Lower horizontal to horizontal center of panel
Upper horizontal to top of the panel
Check Panel Pivot
Use Centering Tool, line centered in panel, not on a diagonal
Check Panel Centering
Panel level to horizontal, use adjustment set screws to adjust
Check Beam Collimator Alignment
Center Collimator, Position A (right offset), Position B (left offset)
Upper Collimator
Perform Gain and Air Normalization Calibration
Nothing in the field of View
Perform Geometric Calibration
Use Geometric Phantom
Perform QA Procedures
QA Phantom Tests (line pairs, HU)
Water Phantom Tests (HU uniformity)
Replacement Parts:
Part Description
Part Number
Gantry Belt
200104
Rotation Motor
100109
Содержание pedCAT
Страница 32: ...31 0473 90001 Rev 07 26 13 Rear Connector Panel Voltage Nameplate location System Label location ...
Страница 34: ...33 0473 90001 Rev 07 26 13 Indicator Panel on machine Patient Platform label Patient Seat label ...
Страница 35: ...34 0473 90001 Rev 07 26 13 Handle Bar Label Cleaning Instructions Label Pinch Point Label Laser Label ...
Страница 37: ...36 0473 90001 Rev 07 26 13 Tube Head Label Beam Limiter Label X Ray Power Supply label ...
Страница 40: ...39 0473 90001 Rev 07 26 13 Laser Lights Button ...
Страница 93: ...92 0473 90001 Rev 07 26 13 Beam Path and Angulation ...
Страница 97: ...96 0473 90001 Rev 07 26 13 ...
Страница 144: ...143 0473 90001 Rev 07 26 13 ...
Страница 168: ...167 0473 90001 Rev 07 26 13 ...