13. To switch between the Axial images and the 3 static 3D images, click on the
RECON/Overview button.
14. Once certain the image looks like the anatomy desired was done so acceptably, click on
the Accept/Next button. If the scan needs to be retaken, click on the “Retake” button to
go back to the Worklist.
NOTE: If there is any indication of vibration to the system, or malfunction to the
system, or computer crash during a scan, please contact CurveBeam Technical Support
for assistance. If there is a failed procedure, turn the machine off by the Emergency
Stop button, following the below procedure.
Procedure for Emergency Removal of a Patient:
The system has undergone extensive testing of the mechanically, electrically and software
performance, but if an unexpected occurrence is observed and/or the software locks up during a scan
or an emergency arises where it becomes necessary to interrupt a scan and/or remove a patient from
the system before a scan is completed, please follow these steps:
1. Press the EMERGENCY STOP button. This will halt the X-ray as well as the motors to the
machine functions and all power to the machine. The message below will display on screen
and will terminate the “CB Scanning Device” Acquisition software:
2. Open the Gate doors manually with your hands by using the slots at the top of each door to
grasp and slide to an open position.
3. Carefully assist the patient to step out of the scan platform area.
4. Reset the machine: Close the PedCAT Acquisition, “CB Scanning Device” software (if not
already). Release the E-stop that was engaged by pressing in and turning the knob to the
right until it pops up. Then turn the machine power back ON at the Main Circuit Breaker. Wait
2 minutes, then re-launch the PedCAT Acquisition “CB Scanning Device” software. Now the
system can be operated again as expected
Содержание pedCAT
Страница 32: ...31 0473 90001 Rev 07 26 13 Rear Connector Panel Voltage Nameplate location System Label location ...
Страница 34: ...33 0473 90001 Rev 07 26 13 Indicator Panel on machine Patient Platform label Patient Seat label ...
Страница 35: ...34 0473 90001 Rev 07 26 13 Handle Bar Label Cleaning Instructions Label Pinch Point Label Laser Label ...
Страница 37: ...36 0473 90001 Rev 07 26 13 Tube Head Label Beam Limiter Label X Ray Power Supply label ...
Страница 40: ...39 0473 90001 Rev 07 26 13 Laser Lights Button ...
Страница 93: ...92 0473 90001 Rev 07 26 13 Beam Path and Angulation ...
Страница 97: ...96 0473 90001 Rev 07 26 13 ...
Страница 144: ...143 0473 90001 Rev 07 26 13 ...
Страница 168: ...167 0473 90001 Rev 07 26 13 ...