74
0473
90001
Rev
07.26.13
Perform the SECOND Scan
(Large FOV) instructions by starting the Geometric
Calibration again using the Large FOV Scan.
Only perform the following steps after both Medium and Large FOV scans have
been completed.
Next the Geometry files that are used by the Recon Manager will need to be copied
and then replaced with the newly created and edited file from above. To do this,
open the “Recon Manager Runtime” folder on the desktop. From here, there are two
folders that will be accessed. The Configuration folder and the ConfigurationPass2
folder. In both of these two folders, locate the file “Geometry_768.xml” and rename it
to “Geometry_768_YYYY_MM_DD.xml” where YYYY is the current year and
MM_DD represents the current (today’s) month and date. Then locate the file
“Geometry_1368.xml” and rename it to “Geometry_1368_YYYY_MM_DD.xml” where
YYYY is the current year and MM_DD represents the current (today’s) month and
date. This will need to be done TWICE, once for the Geometry_768.xml and
Geometry_1368.xml files in the Configuration folder and once for the same files in
the ConfigurationPass2 folder.
Navigate back to the “GeoCal Files” folder that has the newly created and edited
Geometry files. Copy both the Geometry_768.xml and the Geometry_1368.xml files.
Navigate on the desktop to the “Recon Manager Runtime” folder. The copy of the
geometry files will now need to be placed in BOTH the Configuration folder and the
ConfigurationPass2 folder inside the “Recon Manager Runtime” folder. So navigate
to each folder and right click and select “Paste” to paste a copy of the geometry files
in both the Configuration folder and the ConfigurationPass2 folder.
Close the “Recon Manager Runtime” folder.
Close the “GeoCal Files” folder.
Start the Recon Manager Program.
Close the Recon VM window.
Place the Patient Platform in the scanner.
All calibrations are complete.
Содержание pedCAT
Страница 32: ...31 0473 90001 Rev 07 26 13 Rear Connector Panel Voltage Nameplate location System Label location ...
Страница 34: ...33 0473 90001 Rev 07 26 13 Indicator Panel on machine Patient Platform label Patient Seat label ...
Страница 35: ...34 0473 90001 Rev 07 26 13 Handle Bar Label Cleaning Instructions Label Pinch Point Label Laser Label ...
Страница 37: ...36 0473 90001 Rev 07 26 13 Tube Head Label Beam Limiter Label X Ray Power Supply label ...
Страница 40: ...39 0473 90001 Rev 07 26 13 Laser Lights Button ...
Страница 93: ...92 0473 90001 Rev 07 26 13 Beam Path and Angulation ...
Страница 97: ...96 0473 90001 Rev 07 26 13 ...
Страница 144: ...143 0473 90001 Rev 07 26 13 ...
Страница 168: ...167 0473 90001 Rev 07 26 13 ...