Temperature Monitoring Failures:
The system is designed to send a Failure message to
the user if the internal temperature of the machine is 20° C or lower OR 60° C or higher.
This Failure message would display in the “CB Scanning Device” Acquisition software at
Startup and just before a scan is captured. If this condition occurs the error message will
force the acquisition program to terminate and disallow any scanning.
In the event of such Failure message, the user should re-start the system via the Circuit
Breaker ON/OFF switch and then re-launch the “CB Scanning Device” Acquisition software.
If the message still displays, the user should contact CurveBeam technical support.
The Failure Message reads:
Startup:
Note that there is an Internal Temperature gage in the Startup Screen. If in the
Startup Screen and the temperature monitor has detected temperature out of range, the
failure message will display during the Startup sequence.
Internal
Temperature
Gage
Содержание pedCAT
Страница 32: ...31 0473 90001 Rev 07 26 13 Rear Connector Panel Voltage Nameplate location System Label location ...
Страница 34: ...33 0473 90001 Rev 07 26 13 Indicator Panel on machine Patient Platform label Patient Seat label ...
Страница 35: ...34 0473 90001 Rev 07 26 13 Handle Bar Label Cleaning Instructions Label Pinch Point Label Laser Label ...
Страница 37: ...36 0473 90001 Rev 07 26 13 Tube Head Label Beam Limiter Label X Ray Power Supply label ...
Страница 40: ...39 0473 90001 Rev 07 26 13 Laser Lights Button ...
Страница 93: ...92 0473 90001 Rev 07 26 13 Beam Path and Angulation ...
Страница 97: ...96 0473 90001 Rev 07 26 13 ...
Страница 144: ...143 0473 90001 Rev 07 26 13 ...
Страница 168: ...167 0473 90001 Rev 07 26 13 ...