There will be a long delay between some of the calibration steps. There is a
countdown box that will display the seconds until the panel is ready and then the
User Instructions will be updated as to what to do next.
The system will perform the 120kVp calibrations and will require x-ray to be fired
THREE times. The scanner will reposition itself and then perform the 100kVp
calibrations and again fire x-ray THREE times. There will be a total of SIX times in
which x-ray will be fired during this Panel Air Normalization Calibration.
When the calibration is completed, the Air Frames Creation Calibration window will
close and CB-Tools will be displayed on the screen.
WARNING
The X-ray device may be dangerous to the
Patient and Operator if you do not observe and follow operating
instructions. Do not operate this system unless you have
received training to perform a procedure
Countdown until
panel is ready
Содержание pedCAT
Страница 32: ...31 0473 90001 Rev 07 26 13 Rear Connector Panel Voltage Nameplate location System Label location ...
Страница 34: ...33 0473 90001 Rev 07 26 13 Indicator Panel on machine Patient Platform label Patient Seat label ...
Страница 35: ...34 0473 90001 Rev 07 26 13 Handle Bar Label Cleaning Instructions Label Pinch Point Label Laser Label ...
Страница 37: ...36 0473 90001 Rev 07 26 13 Tube Head Label Beam Limiter Label X Ray Power Supply label ...
Страница 40: ...39 0473 90001 Rev 07 26 13 Laser Lights Button ...
Страница 93: ...92 0473 90001 Rev 07 26 13 Beam Path and Angulation ...
Страница 97: ...96 0473 90001 Rev 07 26 13 ...
Страница 144: ...143 0473 90001 Rev 07 26 13 ...
Страница 168: ...167 0473 90001 Rev 07 26 13 ...