23
0473
90001
Rev
07.26.13
proper operation. If the warning system components are found functional
then contact CurveBeam Technical Support.
DOOR INTERLOCK: Next, if you have a Door Interlock, first close the X-
ray Room Door. Click on the DOOR INTERLOCK TEST button. If
functioning properly, then a message will appear indicating so. Then
Open the X-Ray Room Door and click on the DOOR INTERLOCK TEST
button again. The message should indicate that the Door is opened and
no x-ray can be taken at this time.
EXPOSURE: This will check the Exposure Switch, audible exposure
sound and both visual X-Ray ON indicator lights. Click on the
EXPOSURE TEST button. An instruction should display asking to “Push
and hold down the exposure button”. Hold down the exposure switch for
the duration of the exposure, which should be approx. 5 seconds.
Observe for the audible sound and that the X-ray ON lights on the
machine and the operator’s control box are illuminated in Amber color
during this exposure.
EMERGENCY STOP: Exit the CB-Tools program by clicking on the X in
the upper right corner. With the machine Main Circuit Breaker in the ON
position, Power Light ON, press the EMERGENCY STOP button. The
Main Circuit Breaker should trip. Release the EMERGENCY STOP by
turning the knob to the right until it pops up.
Turn the Main Circuit Breaker to
the ON position and re-launch the PedCAT ACQ Software.
Planned Maintenance – Quarterly Schedule:
The system requires quarterly maintenance check for Panel Calibration. This can be
performed by the end user with CurveBeam Technical Support if necessary..
1. The procedure for the panel calibration is in the section under
Calibrations:
Panel for
Gain & Air Calibration
.
Advice: Please refer to that section for instructions on performing this
procedure. (Chapter 4: Calibration & QA Procedures).
If panel calibrations are not routinely performed, there may be suboptimal image
quality. The scan results may have symptoms of artifacts commonly referred to
as “circle or ring artifacts”. Below are samples of circle artifacts in scan results.
If these are observed than a panel calibration should be performed.
Advice: Please refer to Chapter 4 and also contact CurveBeam Technical
Support.
Содержание pedCAT
Страница 32: ...31 0473 90001 Rev 07 26 13 Rear Connector Panel Voltage Nameplate location System Label location ...
Страница 34: ...33 0473 90001 Rev 07 26 13 Indicator Panel on machine Patient Platform label Patient Seat label ...
Страница 35: ...34 0473 90001 Rev 07 26 13 Handle Bar Label Cleaning Instructions Label Pinch Point Label Laser Label ...
Страница 37: ...36 0473 90001 Rev 07 26 13 Tube Head Label Beam Limiter Label X Ray Power Supply label ...
Страница 40: ...39 0473 90001 Rev 07 26 13 Laser Lights Button ...
Страница 93: ...92 0473 90001 Rev 07 26 13 Beam Path and Angulation ...
Страница 97: ...96 0473 90001 Rev 07 26 13 ...
Страница 144: ...143 0473 90001 Rev 07 26 13 ...
Страница 168: ...167 0473 90001 Rev 07 26 13 ...