When prompted, Press and Hold the Scan button to fire the x-ray.
Once the x-ray has completed, release the scan button. The image on the screen
will look similar to the below image:
Observe the values in the yellow boxes.
The left column values for both “C-B (horiz / vert)” and “A – C (horiz / vert)” values
should be 600 +/- 0.5 pixels. The “A – B (horiz / vert)” values in the left column
WARNING
The X-ray device may be dangerous to the
Patient and Operator if you do not observe and follow operating
instructions. Do not operate this system unless you have
received training to perform a procedure