152
0473
90001
Rev
07.26.13
11. Check Laser Light Alignment:
There are 3 laser lights, one vertical in center of rotation, 2 horizontal at 10cm height and 20
cm height.
Vertical laser light: this should be aligned to the platform center screw as depicted in the
below image. If not, adjust it via the adjustment screw on the vertical laser light mechanism
to achieve proper location, circled in the below image.
Horizontal Laser Lights: The lower light should be 10 cm above the platform. The upper
laser light should be 20 cm above the platform. If not, adjust each adjustment screw on the
horizontal laser light mechanisms to achieve proper location.
Содержание pedCAT
Страница 32: ...31 0473 90001 Rev 07 26 13 Rear Connector Panel Voltage Nameplate location System Label location ...
Страница 34: ...33 0473 90001 Rev 07 26 13 Indicator Panel on machine Patient Platform label Patient Seat label ...
Страница 35: ...34 0473 90001 Rev 07 26 13 Handle Bar Label Cleaning Instructions Label Pinch Point Label Laser Label ...
Страница 37: ...36 0473 90001 Rev 07 26 13 Tube Head Label Beam Limiter Label X Ray Power Supply label ...
Страница 40: ...39 0473 90001 Rev 07 26 13 Laser Lights Button ...
Страница 93: ...92 0473 90001 Rev 07 26 13 Beam Path and Angulation ...
Страница 97: ...96 0473 90001 Rev 07 26 13 ...
Страница 144: ...143 0473 90001 Rev 07 26 13 ...
Страница 168: ...167 0473 90001 Rev 07 26 13 ...