78
0473
90001
Rev
07.26.13
2. Place the water phantom on the platform and ensure it is centered using the laser lights
and circles on the patient platform.
3. Acquire a new scan of the water phantom using a
Medium Field (120 kVp) scan
option
.
Advice: Please refer to the Operations: Acquisition section of the manual, Chapter
6, for instructions on acquiring the scan.
4. Once the data is captured, it can be evaluated in the CubeVue, viewing software. Open
the dataset to the Review/MPR Tab’s Axial View. Apply the Sharp Filter and change the
slab thickness on the axial window to the smallest possible slab size.
5.
Noise Level Test:
from the axial view, use the large Hounsfield circular measurements
from within CubeVue and move them so that one is in the center of the water.
The Value should fall in the below range:
Water: -150 to 150