Once the 2 screws are removed, pull the Power Panel out to its operational position and
secure into place with the 4 mounting screws.
4. At the location of placement, set the machine down on its base.
If being placed with the rear to the wall, be sure to leave some room in the back in order
to plug in all cables and for securing the rear cover before placing the machine into its
final spot.
There are a total of 4 adjustment “feet” to the machine, one at each corner. There is
also a center foot that has no adjustment. The machine should sit on the center foot and
lean back onto the 2 back feet. The 2 front feet should not yet be touching the floor at
this time. They will be adjusted once the machine leveling is complete.
Front
leveling
feet.