The Scan Button to start the exposure is mounted on the Operator control box.
4. Deliver the Patient Scan Instructions to the patient.
5. Now Push & Hold the scan button down. During exposure an audible signal is
generated by the machine, and the visual X-ray ON indicator lights will be illuminated in
Amber color. The visual X-ray ON indicators are on the machine, the operator control
box (above) and in PedCAT Acquisition “CB Scanning Device” software. The Operator
should hold the exposure switch for the duration of the exposure as indicated by sound
and lights.
Software
Machine
Exposure Control hand held:
Hand held with coil cord for up to
10 feet away. Can be removed
from the holster.
Scan Button
for initiating the scan.
Must be held down for the duration
of the capture.
Emergency Stop Button
Press down if the exposure needs
to be stopped. This will seize
exposure, and motors.
To Reset the button, turn it to the
right so it pops out.
Status Indicator Lights:
For Power ON, Exposure Ready,
Exposure ON, Fault.
Operator
Control
Box
Содержание pedCAT
Страница 32: ...31 0473 90001 Rev 07 26 13 Rear Connector Panel Voltage Nameplate location System Label location ...
Страница 34: ...33 0473 90001 Rev 07 26 13 Indicator Panel on machine Patient Platform label Patient Seat label ...
Страница 35: ...34 0473 90001 Rev 07 26 13 Handle Bar Label Cleaning Instructions Label Pinch Point Label Laser Label ...
Страница 37: ...36 0473 90001 Rev 07 26 13 Tube Head Label Beam Limiter Label X Ray Power Supply label ...
Страница 40: ...39 0473 90001 Rev 07 26 13 Laser Lights Button ...
Страница 93: ...92 0473 90001 Rev 07 26 13 Beam Path and Angulation ...
Страница 97: ...96 0473 90001 Rev 07 26 13 ...
Страница 144: ...143 0473 90001 Rev 07 26 13 ...
Страница 168: ...167 0473 90001 Rev 07 26 13 ...