07.26.13
5. Panel Gain Calibration:
Remove all objects from the Field of View in the scanner.
Open the CB-Tools by using the shortcut on the desktop:
Select the “Gain Calibration” tab in CB-Tools. Then select “OK” under Run Tool.
When the program is launched the Panel Gain Calibration window will open, with
CB-Tools still visible in the background. Follow the User Instructions for this
calibration. When the User Instruction state to “Select the ‘Start’ button to continue.”
select the “Start Cal” button as shown in the image below.
When going through the instructions, the program will require x-ray to be fired.
There will be a prompt to hold the Scan button, at which point x-ray will be fired. It
Run Tool – OK button
Gain Calibration Tab
User Instructions to
select the “Start” button
Start Cal Button
Use this button when
requested to select
“Start” button