Cleaning:
Routinely and after each patient scan, clean and disinfect all items which come in
contact with the patient.
Cleaning the equipment frequently, especially if corroding chemicals are present is a
function of the Operator. Unless otherwise instructed, use a cloth
moistened
with warm
water and mild soap. Do not use strong cleaners and solvents as these may damage
the finish. Be careful when cleaning to avoid liquid leaking underneath the Platform Area
and into the Gantry. DO NOT USE WATER ON THE IMAGE RECEPTOR, this should
only be dusted with a dry cloth.
For disinfecting, use Biocide ® from Biotrol International or equivalent cleaner and
disinfectant. Biocide® is an Iodophor formulation that kills HIV, Tuberculosis and Polio
in 10 minutes. Moisten a cloth with the disinfectant and then wipe down the area. Do
not spray disinfectant directly onto the equipment.
Preventive Maintenance Schedule - for Owner / User:
Daily:
Routine Dusting - all surfaces.
Clean and Disinfect all items which come in contact with the patient after each
patient scan.
Monthly
:
Clean/Disinfect all surfaces, check for failed/faulty indicator lights.
Quarterly:
Perform Panel Calibration (Gain, Air)
Yearly:
Check for satisfactory image quality.
Содержание pedCAT
Страница 32: ...31 0473 90001 Rev 07 26 13 Rear Connector Panel Voltage Nameplate location System Label location ...
Страница 34: ...33 0473 90001 Rev 07 26 13 Indicator Panel on machine Patient Platform label Patient Seat label ...
Страница 35: ...34 0473 90001 Rev 07 26 13 Handle Bar Label Cleaning Instructions Label Pinch Point Label Laser Label ...
Страница 37: ...36 0473 90001 Rev 07 26 13 Tube Head Label Beam Limiter Label X Ray Power Supply label ...
Страница 40: ...39 0473 90001 Rev 07 26 13 Laser Lights Button ...
Страница 93: ...92 0473 90001 Rev 07 26 13 Beam Path and Angulation ...
Страница 97: ...96 0473 90001 Rev 07 26 13 ...
Страница 144: ...143 0473 90001 Rev 07 26 13 ...
Страница 168: ...167 0473 90001 Rev 07 26 13 ...