61
0473
90001
Rev
07.26.13
Once open, the Input Folder will need to be set to: Y:\Airframes\768x768x100
To do this, select “Browse” on the Input Folder line. Then navigate to the “Y” drive
and the Airframes folder. Click on the 768x768x100 folder. Then click on OK.
The output folder does not get changed and should remain at:
C:\Users\cbadmin\Desktop\ReconManager Runtime\Normalization
Please note, if this is the first time the server is running this procedure, this Output
folder will need to be set. But once set, the folder should remain as specified above
for any future times through this process.
Enter the following values:
KV: 100
Frame Width: 768
Frame Height: 768
# of frames: 128
Click Browse to
search for Input
Folder
Locate Input
Folder
Click OK when Input
Folder location is
selected